Print Options

Card layout: ?

← Back to notecard set|Easy Notecards home page

Instructions for Side by Side Printing
  1. Print the notecards
  2. Fold each page in half along the solid vertical line
  3. Cut out the notecards by cutting along each horizontal dotted line
  4. Optional: Glue, tape or staple the ends of each notecard together
  1. Verify Front of pages is selected for Viewing and print the front of the notecards
  2. Select Back of pages for Viewing and print the back of the notecards
    NOTE: Since the back of the pages are printed in reverse order (last page is printed first), keep the pages in the same order as they were after Step 1. Also, be sure to feed the pages in the same direction as you did in Step 1.
  3. Cut out the notecards by cutting along each horizontal and vertical dotted line
To print: Ctrl+PPrint as a list

53 notecards = 14 pages (4 cards per page)

Viewing:

QUIZ 1 extra

front 1

A plasma sample is hemolyzed and turbid. What
is required to perform a sample blank in order
to correct the measurement for the intrinsic
absorbance of the sample when performing a
spectrophotometric assay?
A. Substitute deionized water for the sample
B. Dilute the sample 1:2 with a standard of
known concentration
C. Substitute saline for the reagent
D. Use a larger volume of the sample

back 1

C. Substitute saline for the reagent

front 2

Which instrument requires a highly regulated
DC power supply?
A. A spectrophotometer with a barrier layer cell
B. A colorimeter with multilayer interference
filters
C. A spectrophotometer with a photomultiplier
tube
D. A densitometer with a photodiode detector

back 2

C. A spectrophotometer with a photomultiplier
tube

front 3

Which statement regarding reflectometry is true?
A. The relation between reflectance density and
concentration is linear
B. Single-point calibration can be used to determine
concentration
C. 100% reflectance is set with an opaque film
called a white reference
D. The diode array is the photodetector of choice

back 3

C. 100% reflectance is set with an opaque film
called a white reference

front 4

Bichromatic measurement of absorbance can
correct for interfering substances if:
A. The contribution of the interferent to absorbance
is the same at both wavelengths
B. Both wavelengths pass through the sample
simultaneously
C. The side band is a harmonic of the primary
wavelength
D. The chromogen has the same absorbance at both
wavelengths

back 4

A. The contribution of the interferent to absorbance
is the same at both wavelengths

front 5

Which instrument requires a primary and
secondary monochromator?
A. Spectrophotometer
B. Atomic absorption spectrophotometer
C. Fluorometer
D. Nephelometer

back 5

C. Fluorometer

front 6

Which of the following statements about
fluorometry is accurate?
A. Fluorometry is less sensitive than
spectrophotometry
B. Fluorometry is less specific than
spectrophotometry
C. Unsaturated cyclic molecules are often
fluorescent
D. Fluorescence is directly proportional to
temperature

back 6

C. Unsaturated cyclic molecules are often
fluorescent

front 7

Which of the following components is not needed
in a chemiluminescent immunoassay analyzer?
A. Source lamp
B. Monochromator
C. Photodetector
D. Wash station

back 7

A. Source lamp

front 8

Which substance is used to generate the light
signal in electrochemiluminescence?
A. Acridinium
B. Luminol
C. Dioxetane phosphate
D. Ruthenium

back 8

D. Ruthenium

front 9

Light scattering when the wavelength is greater
than 10 times the particle diameter is described by:
A. Rayleigh’s law
B. The Beer–Lambert law
C. Mie’s law
D. The Rayleigh–Debye law

back 9

A. Rayleigh’s law

front 10

Which statement regarding nephelometry is true?
A. Nephelometry is less sensitive than absorption
spectrophotometry
B. Nephelometry follows Beer’s law
C. The optical design is identical to a turbidimeter
except that a HeNe laser light source is used
D. The detector response is directly proportional to
concentration

back 10

D. The detector response is directly proportional to
concentration

front 11

The purpose of the nebulizer in an atomic
absorption spectrophotometer that uses
a flame is to:
A. Convert ions to atoms
B. Cause ejection of an outer shell electron
C. Reduce evaporation of the sample
D. Burn off organic impurities

back 11

A. Convert ions to atoms

front 12

A flameless atomic absorption spectrophotometer
dehydrates and atomizes a sample using:
A. A graphite capillary furnace
B. An electron gun
C. A thermoelectric semiconductor
D. A thermospray platform

back 12

A. A graphite capillary furnace

front 13

When measuring lead in whole blood using atomic
absorption spectrophotometry, what reagent is
required to obtain the needed sensitivity and
precision?
A. Lanthanum
B. Lithium
C. Triton X-100
D. Chloride

back 13

C. Triton X-100

front 14

Interference in atomic absorption
spectrophotometry caused by differences
in viscosity is called:
A. Absorption interference
B. Matrix effect
C. Ionization interference
D. Quenching

back 14

B. Matrix effect

front 15

All of the following are required when
measuring magnesium by atomic absorption
spectrophotometry except:
A. A hollow cathode lamp with a magnesium
cathode
B. A chopper to prevent optical interference from
magnesium emission
C. A monochromator to isolate the magnesium
emission line at 285 nm
D. A 285-nm reference beam to correct for
background absorption

back 15

D. A 285-nm reference beam to correct for
background absorption

front 16

When measuring calcium by atomic absorption
spectrophotometry, which is required?
A. An organic extraction reagent to deconjugate
calcium from protein
B. An internal standard
C. A magnesium chelator
D. Lanthanum oxide to chelate phosphates

back 16

D. Lanthanum oxide to chelate phosphates

front 17

Select the equation describing the potential that
develops at the surface of an ion-selective
electrode.
A. van Deemter equation
B. van Slyke equation
C. Nernst equation
D. Henderson–Hasselbalch equation

back 17

A. van Deemter equation

front 18

The reference potential of a silver–silver chloride
electrode is determined by the:
A. Concentration of the potassium chloride filling
solution
B. Surface area of the electrode
C. Activity of total anion in the paste covering the
electrode
D. The concentration of silver in the paste covering
the electrode

back 18

A. Concentration of the potassium chloride filling
solution

front 19

The term RT/nF in the Nernst equation
defines the:
A. Potential at the ion-selective membrane
B. Slope of the electrode
C. Decomposition potential
D. Isopotential point of the electrode

back 19

B. Slope of the electrode

front 20

The ion-selective membrane used to measure
potassium is made of:
A. High-borosilicate glass membrane
B. Polyvinyl chloride dioctylphenyl phosphonate
ion exchanger
C. Valinomycin gel
D. Calomel

back 20

C. Valinomycin gel

front 21

The response of a sodium electrode to a 10-fold
increase in sodium concentration should be:
A. A 10-fold drop in potential
B. An increase in potential of approximately 60 mV
C. An increase in potential of approximately 10 mV
D. A decrease in potential of approximately 10 mV

back 21

B. An increase in potential of approximately 60 mV

front 22

The response of a sodium electrode to a 10-fold
increase in sodium concentration should be:
A. A 10-fold drop in potential
B. An increase in potential of approximately 60 mV
C. An increase in potential of approximately 10 mV
D. A decrease in potential of approximately 10 mV

back 22

B. An increase in potential of approximately 60 mV

front 23

Which of the electrodes below is a current producing (amperometric) rather than a voltage-producing (potentiometric) electrode?
A. Clark electrode
B. Severinghaus electrode
C. pH electrode
D. Ionized calcium electrode

back 23

A. Clark electrode

front 24

Which of the following would cause a “response”
error from an ion-selective electrode for sodium
when measuring serum but not the calibrator?
A. Interference from other electrolytes
B. Protein coating the ion-selective membrane
C. An overrange in sodium concentration
D. Protein binding to sodium ions

back 24

B. Protein coating the ion-selective membrane

front 25

In polarography, the voltage needed to cause
depolarization of the cathode is called the:
A. Half-wave potential
B. Isopotential point
C. Decomposition potential
D. Polarization potential

back 25

B. Isopotential point

front 26

Persistent noise from an ion-selective electrode is
most often caused by:
A. Contamination of sample
B. Blocked junction at the salt bridge
C. Overrange from high concentration
D. Improper calibration

back 26

B. Blocked junction at the salt bridge

front 27

Which element is reduced at the cathode of a
Clark polarographic electrode?
A. Silver
B. Oxygen
C. Chloride
D. Potassium

back 27

B. Oxygen

front 28

Which of the following statements accurately
characterizes the coulometric titration of chloride?
A. The indicator electrodes generate voltage
B. Constant current must be present across the
generator electrodes
C. Silver ions are formed at the generator cathode
D. Chloride concentration is inversely proportional
to titration time

back 28

B. Constant current must be present across the
generator electrodes

front 29

In the coulometric chloride titration:
A. Acetic acid in the titrating solution furnishes
the counter ion for reduction
B. The endpoint is detected by amperometry
C. The titrating reagent contains a phosphate buffer
to keep pH constant
D. Nitric acid (HNO3

back 29

B. The endpoint is detected by amperometry

front 30

Which of the following compounds can interfere
with the coulometric chloride assay?
A. Bromide
B. Ascorbate
C. Acetoacetate
D. Nitrate

back 30

A. Bromide

front 31

All of the following compounds contribute to the
osmolality of plasma except:
A. Lipids
B. Creatinine
C. Drug metabolites
D. Glucose

back 31

A. Lipids

front 32

One mole per kilogram H2O of any solute will
cause all of the following except:
A. Lower the freezing point by 1.86°C
B. Raise vapor pressure by 0.3 mm Hg
C. Raise the boiling point by 0.52°C
D. Raise osmotic pressure by 22.4 atm

back 32

B. Raise vapor pressure by 0.3 mm Hg

front 33

What component of a freezing point osmometer
measures the sample temperature?
A. Thermistor
B. Thermocouple
C. Capacitor
D. Electrode

back 33

B. Thermocouple

front 34

What type of measuring circuit is used in a
freezing point osmometer?
A. Electrometer
B. Potentiometer
C. Wheatstone bridge
D. Thermal conductivity bridge

back 34

C. Wheatstone bridge

front 35

What type of detector is used in high-performance
liquid chromatography with electrochemical
detection (HPLC–ECD)?
A. Calomel electrode
B. Conductivity electrode
C. Glassy carbon electrode
D. Polarographic electrode

back 35

C. Glassy carbon electrode

front 36

In gas chromatography, the elution order of
volatiles is usually based upon the:
A. Boiling point
B. Molecular size
C. Carbon content
D. Polarity

back 36

A. Boiling point

front 37

Select the chemical that is used in most HPLC
procedures to decrease solvent polarity.
A. Hexane
B. Nonane
C. Chloroform
D. Acetonitrile

back 37

D. Acetonitrile

front 38

In thin-layer chromatography (TLC), the distance
the solute migrates divided by the distance the
solvent migrates is the:
A. tR
B. Kd
C. Rf
D. pK

back 38

C. Rf

front 39

Which reagent is used in thin-layer
chromatography (TLC) to extract cocaine
metabolites from urine?
A. Acid and sodium chloride
B. Alkali and organic solvent
C. Chloroform and sodium acetate
D. Neutral solution of ethyl acetate

back 39

B. Alkali and organic solvent

front 40

What is the purpose of an internal standard in
HPLC and GC methods?
A. To compensate for variation in extraction and
injection
B. To correct for background absorbance
C. To compensate for changes in flow rate
D. To correct for coelution of solutes

back 40

A. To compensate for variation in extraction and
injection

front 41

What is the confirmatory method for measuring
drugs of abuse?
A. HPLC
B. Enzyme-multiplied immunoassay technique
(EMIT)
C. Gas chromatography with mass spectroscopy
(GC-MS)
D. TLC

back 41

C. Gas chromatography with mass spectroscopy
(GC-MS)

front 42

What component is used in a GC-MS but not used
in an LC-MS?
A. Electron source
B. Mass filter
C. Detector
D. Vacuum

back 42

A. Electron source

front 43

In addition to velocity, what variable is also
needed to calculate the relative centrifugal force
(g force) of a centrifuge?
A. Head radius
B. Angular velocity coefficient
C. Diameter of the centrifuge tube
D. Ambient temperature in degrees Centigrade

back 43

B. Angular velocity coefficient

front 44

In addition to velocity, what variable is also
needed to calculate the relative centrifugal force
(g force) of a centrifuge?
A. Head radius
B. Angular velocity coefficient
C. Diameter of the centrifuge tube
D. Ambient temperature in degrees Centigrade

back 44

A. Head radius

front 45

When calibrating a semiautomatic pipet that has
a fixed delivery of 10.0 μL using a gravimetric
method, what should be the average weight of
deionized water transferred?
A. 10.0 μg
B. 100.0 μg
C. 1.0 mg
D. 10.0 mg

back 45

D. 10.0 mg

front 46

Which of the following contributes the most to
the serum total CO2?
A. PCO2
B. dCO2
C. HCO3
D. Carbonium ion

back 46

C. HCO3

front 47

Chemiluminescence

back 47

Emission of light by molecules in excited states produced by chemical reactions

front 48

Electrochemistry

back 48

Study of the relationship of electrical potential and chemical changes

front 49

Fluorescence

back 49

Emission of light by a substance after absorption of electromagnetic radiation of a shorter wavelength

front 50

Gas chromatography

back 50

Chromatographic technique used to separate mixtures of compounds that are volatile or can be made volatile

front 51

High-performance liquid chromatography

back 51

Chromatographic technique that uses high pressure and controlled temperature for fast separations of molecules in a mixture

front 52

Ion-selective electrodes

back 52

Electrodes that measure electrical potential produced by the activity of free ions. Each type is designed to be sensitive toward one type of ion

front 53

Osmometry

back 53

Technique used to measure the solute concentration of a solution using one of the four colligative properties, which change in proportion to osmotic pressure